Burns, SP; Delany, AC; Sun, J; Stephens, BB; Oncley, SP; Maclean, GD; Semmer, S R; Schröter, J; Ruppert, J: An Evaluation of Calibration Techniques for In Situ Carbon Dioxide Measurements Using a Programmable Portable Trace-Gas Measuring System, Journal of Atmospheric and Oceanic Technology, 26, 291-316 (2009)
Abstract:
The construction and deployment of a portable trace-gas measurement system (TGaMS) is described. The air-collection system (dubbed HYDRA) collects air samples from 18 different locations and was connected to either one or two LI-COR LI-7000 gas analyzers to measure CO2. An in situ “field calibration” method, that uses four calibration gases with an uncertainty on the order of ±0.1 µmol mol−1 relative to the WMO CO2 mole fraction scale, revealed CO2 output from the LI-7000 had a slightly nonlinear relationship relative to the CO2 concentration of the calibration gases. The sensitivity of the field-calibrated CO2 to different forms of the field-calibration equation is investigated. To evaluate TGaMS performance, CO2 from collocated inlets, portable gas cylinders, and nearby independent CO2 instruments are compared. Results are as follows: 1) CO2 measurements from HYDRA multiple inlets are feasible with a reproducibility of ±0.4 µmol mol−1 (based on the standard deviation of the CO2 difference between collocated inlets when HYDRA was operating with two LI-7000s); 2) CO2 differences among the various field-calibration equations were on the order of ±0.3 µmol mol−1; and 3) comparison of midday hourly CO2 measurements at 30 m AGL between TGaMS and an independent high-accuracy CO2 measurement system (within 300 m of TGaMS) had a median difference and standard deviation of 0.04 ± 0.81 µmol mol−1 over two months.

There are additional file downloads belonging to this publication

Password
This site makes use of cookies More information